Calibration and Auditing Procedures Overview
Discussion of Flow Rate Measurement and General Aspects of PM10/PM2.5 Sampler Calibration
Sampler Calibration
Calibration Apparatus
Precalibration System Check
Rotameter Calibration
Calibration Calculations
  Rotameter Set Point Adjustment Procedure
Sampler Calibration Frequency
MiniVol Flow Rate Performance Audit Using a MiniFlo Transfer Standard
Audit Apparatus
Flow Rate Audit Procedures
Audit Data Calculations
 
» Main MiniVol Page

Calibration and Auditing Procedures
Sampler Calibration Frequency
To ensure accurate measurement of the PM10/PM2.5 concentrations, calibrate the sampler upon installation and then recalibrate it as follows:

  1. At least annually.

  2. After any repairs that might affect sampler calibration.

  3. If the field calibration flow check results exceed QC limits (±7% from the sampler's indicated flow rate).