Rotameter Set Point Adjustment Procedure

  1. Calculate and record the rotameter set point using temperatures and pressures expected on the day(s) sampling is conducted. These values can be seasonal set points or daily settings.


    where:
    Isp = sampler rotameter set point flow rate adjustment
    5.0 = MiniVol design flow rate, L/min
    Pact = ambient barometric pressure, mm Hg
    Tact= ambient temperature, EK.
    Tstd = standard temperature, 298 EK
    Pstd = standard pressure, 760 mm Hg
    mvol = slope of the MiniVol flow rate calibration relationship
    bvol = intercept of the MiniVol flow rate calibration relationship.

    Adjusting the sampler rotameter to seasonal average conditions will help minimize data loss caused by exceeding the MiniVol's design condition specifications.