Rotameter Calibration
Procedures for calibrating the sampler rotameter are as follows:

  1. Install the MiniFlo transfer standard on the inlet tube of the MinVol sampler filter holder.

  2. Turn on the sampler and allow it to warm up to normal operating temperature (at least 2 minutes). The transfer standard must also equilibrate before proceeding with the calibration.

  3. Read and record the following parameters:
    • Ambient temperature (Tact), °K
    • Barometric pressure (Pact), mm Hg
  1. Adjust the flow control until the rotameter displays 6.5 L/min.

  2. Read and record the following parameters:
    • Transfer standard pressure drop
      (H), inches of water
    • Sampler rotameter indication (Qind), liters per minute

  3. Repeat steps 4 and 5 for rotameter settings representing flow rates of 6.0, 5.5, 5.0, 4.5, and 4.0 L/min.