Calibration and Auditing Procedures
Rotameter Calibration
Procedures for calibrating the sampler rotameter are as follows:
-
Install the MiniFlo transfer standard on the inlet tube of the MinVol
sampler filter holder.
-
Turn on the sampler and allow it to warm up to normal operating temperature
(at least 2 minutes). The transfer standard must also equilibrate
before proceeding with the calibration.
- Read and record the following parameters:
- Ambient temperature (Tact), °K
- Barometric
pressure (Pact), mm Hg
-
Adjust the flow control until the rotameter displays 6.5 L/min.
- Read and record the following parameters:
-
Repeat steps 4 and 5 for rotameter settings representing flow rates
of 6.0, 5.5, 5.0, 4.5, and 4.0 L/min.
|